75 Watt Medical Grade Automobile Power Adapter
WWW.ASTRODYNE.COM
FAX: 1-908-353-1021
PHONE: 1-908-353-5700
INDUSTRIES
JEROME
INPUT SPECIFICATIONS
Input Voltage Range
10.8-17.6VDC Continuous
Extended Input Voltage Range for 60 Seconds
10-26VDC DCL100 & DCL200
10-20VDC DCL300
OUTPUT SPECIFICATIONS
Voltage and Current
See Model Charts Page 1
Output Voltage Variance
±5%, includes input & output
condition variation and
component variation
Ripple & Noise (PARD, Note 1)
1% of Vout to 100mV p-p, max.
Current Limit
110% to 160% of Full Load
Non Latching, Auto Recovery
Over Voltage Protection
110% to 135% of Vout
Non Latching, Auto Recovery
Protection
SAFETY & EMC
(DESIGNED TO MEET)
Safety
IEC 60601-1, ANSI/AAMI
ES60601 -1 3rd Ed and
IEC 60601-1-11
Safety Isolation
CF Level Isolation at DC Output
connector terminations
EMC General
IEC 60601-1-2 as applied to
Automotive Environment
Emissions
CISPR11, Radiated Emissions,
Class B
ESD
IEC61000-4-2, ±6kV Direct
Discharge, ±8kV Air Discharge
Radiated Immunity
IEC 61000-4-3, Immunity to radi-
ated RF Electromagnetic field
Conducted Immunity
IEC61000-4-6 Immunity to con-
ducted disturbances induced by
radio frequency fields
EU Automotive Directive 2004/104/EC:
Transient Immunity
Sections 6.8 & 6.9, transient
immunity: ISO 7637-2 test
pulses 1, 2a, 2b, 3a, 3b,
4 & 5 recovery class D
Radiated Emissions
Sections 6.5 & 6.6
Auto Plug Characteristics
UL 20589, pertinent sections
ENVIRONMENTAL SPECIFICATIONS
Operating Temperature
-20°C to 40°C
Derating 20% 40°C to 50°C
70°C for 5 minutes
Storage Temperature
Designed to -25 to +75°C
Operating Altitude
Designed to 0-4,000M
Storage Altitude
Designed to 0-10,000M
Humidity
10% to 95% RH,
Non-Condensing
GENERAL SPECIFICATIONS
Isolation
4500VDC
Efficiency (Note 3)
70-90% Output Dependent
Water Ingress
Per IEC60529, IPX2
MECHANICAL SPECIFICATIONS
Housing Dimensions
4.75” x 3.00” x 1.90”
Excludes Rubber Feet
Input Configuration
Standard Auto Plug
1M Length, 18AWG
Output Configuration
Various; Custom 3’ to 15’
length cords, various output
connectors available
NOTES
1. Measured with a 0.1μF ceramic and a 10μF electrolytic capacitors
across the output terminals. The oscilloscope bandwidth is set at 20MHz
a co-axial cable will be used to measure it. The test condition is maximum
load.



Html Pages

1 2 3


Datasheet Download

Go To PDF Page


Link URL



Does ALLDATASHEET help your business so far?  [ DONATE ] 

About Alldatasheet   |   Advertisement   |   Contact us   |   Privacy Policy   |   Link to Datasheet    |   Link Exchange   |   Manufacturer List
All Rights Reserved©Alldatasheet.com


Mirror Sites
English : Alldatasheet.com  |   English : Alldatasheet.net  |   Chinese : Alldatasheetcn.com  |   German : Alldatasheetde.com  |   Japanese : Alldatasheet.jp
Russian : Alldatasheetru.com  |   Korean : Alldatasheet.co.kr  |   Spanish : Alldatasheet.es  |   French : Alldatasheet.fr  |   Italian : Alldatasheetit.com
Portuguese : Alldatasheetpt.com  |   Polish : Alldatasheet.pl  |   Vietnamese : Alldatasheet.vn
Indian : Alldatasheet.in  |   Mexican : Alldatasheet.com.mx  |   British : Alldatasheet.co.uk  |   New Zealand : Alldatasheet.co.nz
Family Site : ic2ic.com  |   icmetro.com