7 / 8 page ![]() MINT1150 150W Single Output Medical Series www.slpower.com MINT1150 REV 1.4, 13-NOV-12 7 Output Transient Response 50% load step within the regulation limits of minimum and maximum load, dI/dt< 0.2A/μSec. Recovery time not specified as there is no laps in regulation with a 50% Load Step. Maximum voltage deviation is 3%, This test is performed on the MAIN OUTPUT ONLY. 24V OUT, 120VAC, 25% TO 75% LOAD STEP 24V OUT, 240VAC, 25% TO 75% LOAD STEP Output Overload Characteristic Supply shall protect itself against Overload conditions. The Power Supply shall recover from Overload Conditions without operator intervention. 24V OUT, 90VAC 24V OUT, 264VAC Short Circuit Protection Supply shall protect itself against Short Circuit conditions. No damage will occur if the output is shorted. 24V OUT, 90VAC 24V OUT, 264VAC |
|
|
Link URL |
| Does ALLDATASHEET help your business so far? [ DONATE ] |
About Alldatasheet | Advertisement | Contact us | Privacy Policy | Link to Datasheet | Link Exchange | Manufacturer List All Rights Reserved©Alldatasheet.com |
| Russian : Alldatasheetru.com | Korean : Alldatasheet.co.kr | Spanish : Alldatasheet.es | French : Alldatasheet.fr | Italian : Alldatasheetit.com Portuguese : Alldatasheetpt.com | Polish : Alldatasheet.pl | Vietnamese : Alldatasheet.vn Indian : Alldatasheet.in | Mexican : Alldatasheet.com.mx | British : Alldatasheet.co.uk | New Zealand : Alldatasheet.co.nz |
|