42 / 80 page ![]() ADAS1000/ADAS1000-1/ADAS1000-2 Data Sheet Rev. A | Page 42 of 80 Pace Amplitude Threshold This register (Address 0x07, see Table 32) can be used to set the minimum valid pace pulse amplitude: PACEAMPTH setting = 16 2 × × GAIN VREF N , (corresponds to a range of 20 µV to 5 mV in the ×1.4 gain setting (GAIN 0)) where: N = 0 to 255 (8 bits), register default N = 0x24, (PACEAMPTH = 706 μV in the ×1.4 gain setting) GAIN = 1.4, 2.1, 2.8, or 4.2 (programmable). VREF = 1.8 V. This level would typically be set to be the minimum expected pace amplitude. For best results under most operating conditions for both biventricular and unipolar pacing, it is suggested to set the pace amplitude threshold to a value of approximately 700 μV to 1 mV. It is not recommended to set the threshold below 250 μV to avoid ambient noise from the patient. The amplitude may need to be adjusted much higher than 1 mV when other medical devices are connected to the patient. Pace Edge Threshold This programmable level (Address 0x0E, see Table 39) is used to find a leading edge, signifying the start of a pace pulse: PACEEDGETH setting = 16 2 × × GAIN VREF N (corresponds to a range of 20 µV to 5 mV in the ×1.4 gain setting) where: If N = 0, PACEEDGETH = PACEAMPTH/2, then N = 0 to 255 (8 bits). GAIN = 1.4, 2.1, 2.8 or 4.2 (programmable). VREF = 1.8 V. Pace Level Threshold This programmable level (Address 0x0F, see Table 40) is used to find the leading edge peak: PACELVLTH setting = 16 2 × × GAIN VREF N , signed (FF = −1, 01 = +1), default = 0 where: N = 0 to 255 (8 bits). GAIN = 1.4, 2.1, 2.8 or 4.2 (programmable). VREF = 1.8 V. Pace Validation Filter 1 This filter is used to reject sub threshold pulses such as minute ventilation (MV) pulses and inductive coupled implantable telemetry systems. It is typically enabled and is controlled via the PACECTL register, Bit 9 (see Table 29) register. Filter 1 applies to all leads enabled for pace detection. Pace Validation Filter 2 This filter is also used to reject sub threshold pulses such as MV pulses and inductive implantable telemetry systems. It is normally enabled and is controlled via the PACECTL register, Bit 10 (see Table 29). Filter 2 applies to all leads enabled for pace detection. Pace Width Filter When enabled, this filter searches for an edge of opposite polarity to the leading edge with a magnitude of at least half the original trigger. The second edge must be between 100 µs to 2 ms from the original edge. When a valid pace width is detected, the width is stored. When disabled only the minimum pulse width of 100 μs is disabled. This filter is controlled by the PACECTL register, Bit 11 (see Table 29). BIVENTRICULAR PACERS As described previously, the pace algorithm expects the pace pulse to be less than 2 ms wide. In a pacer where both ventricles are paced, they can be paced simultaneously. Where they fall within the width and height limits programmed into the algorithm, a valid pace will be flagged, but only one pace pulse may be visible. With the pace width filter enabled, the pace algorithm seeks pace pulse widths within a 100 μs to 2 ms window. Assuming that this filter is enabled and in a scenario where two ventricle pacer pulses fire at slightly different times, resulting in the pulse showing in the lead as one large, wider pulse, a valid pace is flagged so long as the total width does not exceed 2 ms. PACE DETECTION MEASUREMENTS Design verification of the ADAS1000 digital pace algorithm includes detection of a range of simulated pace signals in addition to using the ADAS1000 and evaluation board with one pacemaker device connected to various simulated loads (approximately 200 Ω to over 2 kΩ) and covering the following 4 waveform corners. • Minimum pulse width (100 μs), minimum height (to <300 μV) • Minimum pulse width (100 μs), maximum height (up to 1.0 V) • Maximum pulse width (2 ms), minimum height (to <300 μV) • Maximum pulse width (2 ms), maximum height (up to 1.0 V) These scenarios passed with acceptable results. The use of the ac lead-off function had no obvious impact on the recorded pace height, width, or the ability of the pace detection algorithm to identify a pace pulse. The pace algorithm was also evaluated with the respiration carrier enabled; again, no differences in the threshold or pacer detect were noted from the carrier. |
|
Link URL |
| Does ALLDATASHEET help your business so far? [ DONATE ] |
About Alldatasheet | Advertisement | Contact us | Privacy Policy | Link to Datasheet | Link Exchange | Manufacturer List All Rights Reserved©Alldatasheet.com |
| Russian : Alldatasheetru.com | Korean : Alldatasheet.co.kr | Spanish : Alldatasheet.es | French : Alldatasheet.fr | Italian : Alldatasheetit.com Portuguese : Alldatasheetpt.com | Polish : Alldatasheet.pl | Vietnamese : Alldatasheet.vn Indian : Alldatasheet.in | Mexican : Alldatasheet.com.mx | British : Alldatasheet.co.uk | New Zealand : Alldatasheet.co.nz |
|
Family Site : ic2ic.com |
icmetro.com |